I/O Interconnect Cable for the 4194 or 4195 Network Analyzers

This 50 conductor cable connects the 4194 or 4195 network analyzer display to the test set. Standard length is 18 inches; other lengths available upon request.

NOTICE: DIRTY CONTACTS DROP BITS
The 50-pin ribbon connector contacts are susceptible to contamination that can lead to intermittent instrument operation or loss of functionality. Inspect the 50-pin ribbon connector contacts on the instrument chassis and mating cable for dirt, lint, debris, or other contamination. If found, clean contacts with isopropyl alcohol and blow-dry with clean dry air. Use a lint-free cloth or swab.

Order Part Number: 04194-61602